Gastec 201H 황화물 용액 튜브 (0.02 ~ 0.2mg)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Indicator: 0.02 ~ 0.2 mg 
Measuring Range: 0.02 ~ 0.2 mg 
Sampling Time: 45 seconds per 1 pomp stroke 
Colour Change: White → Brown
Reaction Principle: 
+ Gus generation tube : Sulphides + H2SO4 → H2S
+ Detector tube : H2S + Pb(CH3COO)2 → PbS
Coefficient of Variation: 5% (for 0.02 to 0.2 mg)
Nr. of tests: 10
Option Kit: HYDROTEC-S [No.330]

Catalog

 

Performance
Measuring Range 0.02 to 0.2 mg
Sampling Time 45 seconds per 1 pomp stroke
Detecting Limit 0.002 mg
Colour Change White → Brown
Reaction Principle Gus generation tube : Sulphides + H2SO4 → H2S
Detector tube : H2S + Pb(CH3COO)2 → PbS
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Coefficient of Variation 5% (for 0.02 to 0.2 mg)
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Changes colour by itself to
Chloride ion   No No
Chromate ion   No No
Carbonate ion   No No
Calibration gas generation Static gas dilution method

 

For measuring sulphides in waste water and sludge
HYDROTEC-S
[No.330]

This kit enables quick measurement (approx. 10 min.) of sulphide in bodies of water or in waste water without any special equipment and skills. Take a sample with the gas generator tube, add sulphuric acid and simply measure the amount of hydrogen sulphide generated using the detector tube.
201H / 201L
Measurement procedure
1 The measured out mud and soil is put in the generation tube and 5mL of distilled water are added.
In case of muddy water, it suffices to add a small amount of distilled water (measured out in a graduated cylinder) to the gas generation tube.
Put the cap on the generation tube.
2 Both ends of the detector tube (No.201H or No.201L) are broken off, and one end of the detector tube is connected to the gas generation tube and the other to the gas sampling pump.
3 18N 2 mL of sulphuric acid are added to the gas generation tube.
4 Pull the handle of the gas sampling pump to draw in a sample of the generated hydrogen sulphide.
(By repeating this process several times, all sulphides become hydrogen sulphide several times repeatedly).
The scale in the detector tube is read, and the concentration of the sulphide in the sample is calculated.

 

 