For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 231 Sulphuryl Fluoride SO2F2 검출기 튜브 (1 ~ 20 ppm, Pyrotec Pyrolyzer 860)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Indicator: 1 ~ 20 ppm
Measuring Range: 1 ~ 20 ppm
Sampling Time: 2 minute per pump stroke
Colour Change: Bluish purple→White
Reaction principle: SO2F2 → SO2, SO2 + l2 + H2O → HI + H2SO4
Nr. of tests: 4 tubes

Gas Sampler pump: GV-100S, GV-110S, Pyrotec Pyrolyzer 860

Catalog

Performance
Measuring Range 1 to 20 ppm
Number of Pump Strokes 3
Correction Factor 1
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.4 (n=3)
Colour Change Bluish purple→White
Reaction Principle SO2F2 → SO2
SO2 + l2 + H2O → HI + H2SO4
Coefficient of Variation 10% (for 1 to 20ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Sulphur Dioxide ≧0.04 ppm + White at 0.05 ppm
Carbon Dioxide ≦5 % No No
Chloropicrin(CP) ≦50ppm No No (50ppm)
Methyl Isothiocyanate(MITC) ≦50ppm No No (50ppm)
Low Oxygen environment   + No
Calibration gas generation Vaper pressure method
TLV-TWA TLV-STEL Explosive range
5ppm(2014) 10ppm(2014) --

 

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관 및 개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결