Gastec 5D Passive Dosl Tubes 이산화황 SO2 검지관 (0.2 ~ 100ppm)
Model5D
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능
Indicator: 2~100ppm
Measuring Range: 0.2~100ppm
Sampling Hours: 1~10 hours
Colour Change: Green → Yellow
Reaction Principle: SO2 + BaCl2 + H2O → BaSO3 + 2HCl, HCl + Base → Chloride
Coefficient of Variation: 10% (for 2 to 10 ppm-hr),5% (for 10 to 100 ppm-hr)
Nr. of tests: 10
sulphur dioxide
| NO.5D |
![]() |
Performance
| Measuring Range | 0.2 to 100 ppm |
| Sampling Hours | 1 to 10 hours |
| Detecting Limit | 0.1 ppm (10 hours) |
| Colour Change | Green → Yellow |
| Reaction Principle | SO2 + BaCl2 + H2O → BaSO3 + 2HCl HCl + Base → Chloride |
| Coefficient of Variation | 10% (for 2 to 10 ppm-hr),5% (for 10 to 100 ppm-hr) |
| Shelf Life | 3 Years |
| Corrections for temperature & humidity | Unnecessary |
| Store the tubes at cool and dark place. | |
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Nitrogen dioxide | No | ||
| Hydrogen sulphide | + | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| - | 0.25 ppm (2014) | - |
빠른 지원