For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 60 빠른 측정 감지기 튜브 페놀 C6H5OH (0.4 ~ 187ppm)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Indicator: 
Measuring Range: 
+ 0.4~1ppm, Number of Pump Strokes 4
+ 1~25ppm, Number of Pump Strokes 2
+ 25~62.5ppm, Number of Pump Strokes 1
+ 62.5ppm~187ppm, Number of Pump Strokes 1/2
Sampling Time: 1.5 minute per pump stroke
Colour Change: Pale yellow → Gray
Reaction Principle: C6H5OH + Ce(NO3)6^2- → C6H5OCe(NO3)5N
Coefficient of Variation: 15% (for 1 to 5 ppm), 10% (for 5 to 25 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range
0.4 to
1 ppm
1 to
25 ppm
25 to
62.5 ppm
62.5 to
187 ppm
Number of Pump Strokes
4 2 1 1/2
Correction Factor
0.4 1 2.5 7.5
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.1 ppm (n = 4)
Colour Change Pale yellow → Gray
Reaction Principle C6H5OH + Ce(NO3)62- → C6H5OCe(NO3)5N
Coefficient of Variation 15% (for 1 to 5 ppm), 10% (for 5 to 25 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ammonia 2000 ppm Unclear demarcation White
Amines 2000 ppm Unclear demarcation White
Cresol   + Gray
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Naphthalene by Scale 2 0.5 to 14 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
5 ppm - 3 to 10 %
혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관 및 개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결