Indicator: 5 to 50 ppm
Measuring Range:
+ 2.5 to 5 ppm, Number of Pump Strokes 4
+ 5 to 50 ppm, Number of Pump Strokes 2
+ 50 to 100 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minutes per pump stroke
Colour Change: White → Pale yellow
Reaction Principle: 2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl23H2O+6HCl
Coefficient of Variation: 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
Performance
| Measuring Range |
| 2.5 to 5 ppm |
5 to 50 ppm |
50 to 100 ppm |
|
| Number of Pump Strokes |
|
| Correction Factor |
|
| Sampling Time |
1.5 minutes per pump stroke |
| Detecting Limit |
0.5 ppm (n = 4) |
| Colour Change |
White → Pale yellow |
| Reaction Principle |
2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl2 3H2O+6HCl |
| Coefficient of Variation |
10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm) |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Arsine |
20 ppm |
+ |
Brown |
| Hydrogen selenide |
20 ppm |
+ |
Brown |
| Hydrogen sulphide |
10 ppm |
+ |
Dark brown |
| Calibration gas generation |
High pressure gas cylinder method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 0.3 ppm |
1 ppm |
- |