Indicator: 5 to 120 ppm
Measuring Range:
+ 0.5 to 5 ppm, Number of Pump Strokes 2 to 10
+ 5 to 120 ppm, Number of Pump Strokes 1
Sampling Time: 2 minute per pump stroke
Colour Change: White → Yellow
Reaction Principle: 2C2H5SH + PdSO4 → (C2H5S)2Pd+ H2SO4
Coefficient of Variation: 10% (for 5 to 40 ppm), 5% (for 40 to 120 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 2 minutes per pump stroke | ||
| Detecting Limit | 0.2 ppm (n=10) | ||
| Colour Change | White → Yellow | ||
| Reaction Principle | 2C2H5SH + PdSO4 → (C2H5S)2Pd+ H2SO4 | ||
| Coefficient of Variation | 10% (for 5 to 40 ppm), 5% (for 40 to 120 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity | Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Acetylene, Ethylene | + | Pale brown (whole layer) | |
| Carbon Monoxide | + | Pale brown (whole layer) | |
| Other Mercaptans | + | Yellow | |
| Hydrogen sulphide | + | No |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.5 ppm(2014) | - | 2.8 to 18.2% |
빠른 지원