Gastec 8LL 빠른 측정 감지기 튜브 염소 Cl2 (0.025 ~ 2ppm)
Model8LL
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능
Indicator: 0.1~1ppm
Measuring Range:
+ 0.025 to 0.05 ppm, Number of Pump Strokes 2
+ 0.05 to 1.0 ppm, Number of Pump Strokes 1
+ 1.0 to 2.0 ppm, Number of Pump Strokes 1/2
Sampling Time: 2 minute per pump stroke
Colour Change: White → Pale Green
Reaction Principle: Cl2 + ABTS → Pale green products
Coefficient of Variation: 10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 1.0 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 2 minutes per pump stroke | |||
| Detecting Limit | 0.005 ppm (n=2) | |||
| Colour Change | White → Pale Green | |||
| Reaction Principle | Cl2 + ABTS → Pale green products | |||
| Coefficient of Variation | 10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 1.0 ppm) | |||
| Shelf Life | 1 Year | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Ammonia | No | No | |
| Carbon monoxide, Carbon dioxide | No | No | |
| Hydrogen chloride | No | No | |
| sulphur dioxide | No | No | |
| Nitrogen dioxide, Chlorine dioxide | + | Pale green | |
| Halogens | + | Pale green | |
| Hydrogen sulphide | No | No | |
| Organic vapours | No | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.5 ppm (2014) | 1 ppm (2014) | - |