인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
CALIBRATION
TRACEABILITY 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
Temperature : 18 °C to 28 °C
Humidity: Up to 80 % RH
STORAGE CONDITIONS
Temperature: -20 °C to 50 °C
Humidity: 0 % to 95 % RH (non-condensing)
Altitude: 2000 m
CONTROLLED RANGE
Temperature: 5°C to 55°C
Humidity @ 18°C : 5 % to 95 % RH
Humidity @ 35°C 5 % to 85 % RH
Humidity @ 23°C: 5 % to 95 % RH
Humidity @ 40°C 5 % to 75 % RH
Humidity @ 30°C: 5 % to 90 % RH
Humidity @ 50°C 5 % to 50 % RH
CALIBRATION SYSTEM ACCURACY
Temperature ± 0.12 °C Or Better Typically ± 0.10 °C
Humidity ± 1.00 % RH Or Better Typically ± 0.80 %
Controller type: PID Controller
UNIFORMITY
Temperature: 0.10 °C Or Better Typically ± 0.10 °C
Humidity: 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
Temperature: 0.05 °C
Humidity: 0.15 % RH @ 18 °C to 28 °C
RAMP / SOAK RATE OF CHANGE
Depends on ambient conditions
Temperature
High → Low 1.50 °C / minute (Typical)
Low → High 3.00 °C / minute (Typical)
Humidity
High → Low 5.00 % RH / minute (Typical)
Low → High 10.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition 30 Minutes Maximum
Cold 30 Minutes
RESOLUTION
Display
Temperature: 0.01 °C
Humidity: 0.01 % RH
Dew Point: 0.01 °C (Calculated)
USB
Temperature: 0.01 °C
Humidity: 0.01 % RH