CALIBRATION
TRACEABILITY: 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
Temperature : 18 °C to 28 °C
Humidity: Up to 80 % RH
STORAGE CONDITIONS
Temperature: -20 °C to 50 °C
Humidity: 0 % to 95 % RH (non-condensing)
CONTROLLED RANGE
Temperature : 18 °C to 35 °C @
Ambient Temperature 23 °C
Humidity @ 18°C : 5 % to 75% RH
Humidity @ 23°C : 5 % to 85 % RH
Humidity @ 30°C: 5 % to 80 % RH
CALIBRATION SYSTEM ACCURACY
Temperature: ± 0.10 °C
Humidity: ± 1.00 % RH
Controller type :PID Controller
UNIFORMITY
Temperature : 0.20 °C
Humidity: 0.40 % RH @ 23 °C
STABILITY
Temperature: 0.10 °C
Humidity: 0.15 % RH @ 23 °C
RAMP / SOAK RATE OF CHANGE
*Depends on ambient conditions and desiccant conditions
Temperature
High → Low 1.00 °C / minute (Typical)
Low → High 1.00 °C / minute (Typical)
Humidity
High → Low 3.00 % RH / minute (Typical)
Low → High 3.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition: 30 Minutes Maximum
Cold : 45 to 60 Minutes
RESOLUTION DISPLAY
Temperature 0.1 °C
Humidity 0.1 % RH