Gigahertz-Optik TR-9600-2 Time Data Logger Optometer (100 ns)

ModelTR-9600-2
P/N15295277
원산지Germany
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

-Calibration: Calibration and comparison of the current responsivity in each of amplifier range.

-Measurement range: 

1 µs rise time amplifier: 10 (1 mA/V – 30 nA/V)

100 ns rise time amplifier: 4 (300 µA/V – 10 µA/V)

-Display: LEDs:

POWER: Device switched on, DC-power supply existing

CONNECTED: Remote-Commands received

ERROR: Error, DC-power supply low

-Analog input: BNC-Socket (max. ± 5 mA, max. ± 5 V)

-Bias: -12 V to + 12 V, 10 mA, Shielding BNC-Socket

-Trigger: External trigger Input (TTL/CMOS, positive edge or level)

-Trigger: Trigger Output (CMOS, 4 mA)

-Trigger: Trigger Hysterese: ca. 50 mV

-Measurement range: 10 (three-step,1 mA/V – 30 nA/V)

-ADC: 12 Bit, max. 10 M Samples/s

-Sampling rate: 5 Ranges (10 M Samples/s – 1 k Sample/s)

-Data logger: 

2 M Samples (4 MByte)

-Version: 

TR-9600-2

100 ns version

-Power Supply: 

(6 - 8) V / 1.2 A, DC power plug 5.5 mm / 2.5 mm / 10 mm

AC/DC power supply (7V/1.5 A)

-Temperature range: (5  – 40)°C

-Interface: 

RS232 (75 - 57600Baud, 8 Data-Bit's, 1 Stop-Bit, no Parity)

DSUB9-Socket:

PIN 2: TxD

PIN 3: RxD

PIN 5: GND

PIN 1,4,6 connected

PIN 7,8 connected

IEEE488 (488-1978, AH1, SH1, L4, T4)

-Dimensions: 

w = 28,0 cm

d = 25,2 cm

h = 7,8 cm

-Weight: 2000 g (without battery)

-Warranty: 12 months

-Humidity: <80%, non-condensing