For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

HACH DR4900 Laboratory Spectrophotometer (±1.5 nm (340–900 nm), w/camera)

ModelDR4900
P/NLPV451.99.00121
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Accuracy: ±1.5 nm (340–900 nm)

Altitude: 2000 m maximum

Camera: None

Certifications: CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV

Data Logger: > 10,000 measured values (measured values, date, time, sample ID, user ID)

Dimensions (H × W × D): 226 mm × 255 mm × 344 mm

Display: 10 inch (25.4 cm)

Enclosure Rating: IP20

Indoor/Outdoor: Indoor

Interfaces: USB Type A and Type C (front and back); 1× Ethernet

Light Source: Halogen lamp

Operating Conditions: 10–40°C, maximum 80% relative humidity, non-condensing

Operating Mode: Transmittance (%), absorbance (Abs) or concentration

Optical System: Reference beam, spectral

Overvoltage Category: II

Photometric Accuracy: 5 mAbs at 0.0–0.5 Abs; 1% at 0.50–2.0 Abs

Photometric Linearity: < 0.5% at 2 Abs; ≤ 1% at > 2 Abs with neutral glass at 546 nm

Photometric Measuring Range: ±3.0 Abs (340–900 nm)

Pollution Degree: 2

Power Requirements (Frequency): 50/60 Hz

Power Requirements (Voltage): 100–240 VAC

Power Supply: Benchtop Power Supply

Preprogrammed Methods: > 240

Protection Class: Class I

Sample Cell Compatibility: Rectangular: 10 mm, 50 mm, 1 inch; Round: 13 mm, 16 mm, 1 inch

Scanning Speed: ≥ 8 nm/s (in increments of 1 nm)

Spectral Bandwidth: 5 nm

Storage Conditions: −40 to 60°C, maximum 80% relative humidity, non-condensing

Stray Light: < 0.1% T at 340 nm with NaNO₂

User Programs: > 100 user programs

Wavelength Accuracy: ±1.5 nm (wavelength range 320–900 nm)

Wavelength Calibration: Automatic

Wavelength Range: 320–1100 nm

Wavelength Reproducibility: ±0.1

Wavelength Resolution: 1

Wavelength Selection: Automatic, based on the procedure selected

Weight: 4.8 kg (10.58 lb)

What's Included?: Spectrophotometer DR4900; adapter "A" for 1-inch round cell, 10 mm rectangular cell and Accuvac vials; sample cells (matched pair); light shield; dust cover; power cord with country-specific adapter plug; power adapter; user manual

Datasheet