인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

Mainframe
Microprocessor controlled touch panel: 7”, capacitive
Optical Ethernet Interface for remote control of the generator: optional
Interface for saving reports: USB
External Trigger input/ output: Switch/ 10V
Connector for external safety interlock loop: 24 V=
External red and green warning lamps: 24 V=, 40 mA
Mains power: 230 V, 50/60 Hz
Dimensions of desk top case (W * H * D): 453 * 320 * 520 mm³
Weight: 40 kg
HV-Generator Section
Test voltage, adjustable: 100 - 10,000 V~
Test current: 1 - 100 mA
Test current, adjustable: 2 - 100 mA
Short circuit current (according to VDE 0551): > 200 mA
Test mode, available: turn-off / burn
High voltage output, referred to ground: 4 mm ∅
External measurement output for the test voltage: coaxial, BNC
Tolerance of output voltage amplitude: ± 2% v.E.W.
Rise of sinus voltage ramp, adjustable: 0.1 kV/s - 10 kV/s
Test time, adjustable (1 sec/Step, turn-off & burn): 1 - 100 sec
Test time, adjustable (10 sec/Step, turn-off only): 1 - 1000 sec
Test cabinet, PA 504
Safety interlock connector: break contact, 24V-dc
Red and green warning lights (acc. to VDE 0104): 220V / 10 W
High-voltage output connector: 4 mm ∅
Interconnection cable: 0.5 m long
Dimensions (W * H * D): 555 * 480 * 600 mm³
Test space (W * H * D): 460 * 300 * 550 mm³
Ground rod included
Accessories: Mains cable, Key, Ground rod, Instruction manual.