HIOKI ST5680A DC Hipot Tester (0.010 kV DC~ 8.000 kV, 500 kS/s)
Main Functions
DC hipot test
Insulation resistance test
Breakdown voltage test
Waveform display
Arc discharge detection
Contact check
DC Hipot Test
Output voltage: DC 0.010–8.000 kV (1 V resolution)
Output setting accuracy:
±20 V (below 3 kV, no load)
±0.4% of setting ±8 V (3 kV or higher, no load)
Output/cutoff current: Maximum 20 mA
Current accuracy:
Below 3.00 mA: ±(1.5% of reading + 2 μA)
3.00 mA or higher: ±1.5% of reading
For 0 °C ≤ t* < 18 °C: Base accuracy + base accuracy × [1 + 0.1 × (18 − t*)]
For 28 °C < t* ≤ 40 °C: Base accuracy + base accuracy × [1 + 0.1 × (t* − 28)]
t: Ambient temperature
Maximum resolution: 0.001 μA
Test time: 0.1–999 s or continuous (timer off)
Voltage ramp-up/ramp-down time: Ramp-up: 0.1–300 s; ramp-down: 0.1–300 s; or off
Test modes: W to IR, IR to W, program test
Insulation Resistance Test
Output voltage: DC 10–2,000 V (1 V resolution)
Output setting accuracy: ±20 V (no load)
Resistance display range: 100.0 kΩ–200.0 GΩ (0.01 kΩ resolution)
Guaranteed accuracy range: 100.0 kΩ–99.99 GΩ
Resistance accuracy: See “Insulation resistance measurement accuracy” below for details
Test time: 0.1–999 s or continuous (timer off)
Voltage rise/fall time: Rise: 0.1–300 s; fall: 0.1–300 s; or off
Breakdown Voltage (BDV) Measurement
Test methods: Continuous voltage ramp test or stepped voltage ramp test
Settings: Insulation breakdown voltage (kV), insulation breakdown strength (kV/mm)
Setting details: Start voltage, end voltage, ramp speed, arc detection, electrode distance, upper current limit
Waveform Display
Types: Voltage, current, insulation resistance
Sampling rate: Maximum 500 kS/s
Display length setting: 0.5–128 s (9 selectable settings)
Memory capacity: 512 kilowords
Arc Detection
Detection method: Monitoring fluctuations in the test voltage
Setting: Test voltage variability: 1–50%
Contact Check Function
Detection method: Capacitance measurement
User-definable setting: Capacitance threshold: 1.0–100.0 nF
Memory Functions
Waveform/graph saving: Save to USB memory; supported formats: BMP, PNG, CSV
Panel memory function:
Stores test condition settings internally
DC withstand voltage testing and insulation resistance testing: Up to 64 setting sets for each function
Program testing: Up to 30 programs, with a maximum of 50 steps per program
BDV measurement: Up to 10 setting sets
Data memory function: Stores measured values in the instrument’s internal memory, up to 32,000 values
Judgment Functions
Judgment outputs: PASS, FAIL (UPPER FAIL, LOWER FAIL)
UPPER_FAIL: Measured value > user-defined upper limit
PASS: User-defined upper limit ≥ measured value ≥ user-defined lower limit
LOWER_FAIL: Measured value < user-defined lower limit
List of Major Functions
Interlock: Disables output based on the status of an external device
Auto discharge: Discharges the test object through internal circuitry at the end of the test
Discharge resistance: 716 kΩ
Residual voltage: 30 V or less
Offset cancellation: Measures current flowing along the test path and subtracts it from measurement results
Measurement speed: NORMAL (100 ms), FAST (20 ms), FAST2 (10 ms)
Momentary output: Outputs the test voltage only while the START button is pressed
Command monitor: Displays commands being sent and received on the screen
I/O handler test: Checks whether signals are being properly input and output through the EXT. I/O terminal
Key lock: Prevents changes to test conditions
Self-check: Checks the touch screen, display, LED, instrument memory, and EXT. I/O
Calibration deadline check: Allows a calibration deadline to be set in advance and displays a warning when it has passed
EXT SW: Enables remote operation
Options: Single remote control box 9613, dual remote control box 9614
Basic Specifications
Operating temperature and humidity range: 0–40 °C, 80% RH or less, non-condensing
Compliance standards:
Safety: IEC 61010
EMC: EN 61326
Power supply: 100–240 V AC
Power consumption: Approximately 180 VA
Measurement conditions: 220 V supply voltage, 50/60 Hz supply frequency, DC withstand voltage test mode, 2.5 kV test voltage, and 5 mA load current (500 kΩ load resistance)
Maximum rated power: 800 VA
Interfaces:
Standard: USB, LAN, EXT. I/O
Optional: RS-232C (Z3001), GP-IB (Z3000)
External memory: USB drive
External dimensions (W × H × D): 305 × 142 × 430 mm (excluding protruding parts)
Weight: 10.4 kg ±0.2 kg
인증된 전문가에게 직접 연결