HORIBA HU-200SS Industrial-use MLSS meter (0 ~ 20000 mg/L, including SS-90 sensor)
Measurable Range
Activated sludge (MLSS) : 0 to 20000 mg/L (MLSS)
Kaolin : 0 to 10000 mg/L (white inorganic powder)
Clay (inorganic mud) : 0 to 20000 mg/L (gray inorganic mud)
Transmission Output Setting Range
Activated sludge (MLSS) : Arbitrary setting within the range from 0 to 1000 mg/L to 0 to 20000 mg/L
Kaolin : Arbitrary setting within the range from 0 to 1000 mg/L to 0 to 10000 mg/L
Clay (inorganic mud) : Arbitrary setting within the range from 0 to 1000 mg/L to 0 to 20000 mg/L
Display Resolution
Display resolution : 1 mg/L (0 to 10000 mg/L)
Display resolution : 10 mg/L (10000 to 22000 mg/L)
Performance
Repeatability : Within a reading value ±3% or ±10 mg/L, whichever is larger (sensor connecting sludge measured value)
Transmission Output
Number of output points : 1
Output type : DC 4 mA to 20 mA, input/output insulation type
Load resistance : 900 Ω maximum
Linearity : Within ±0.08 mA (for output only)
Repeatability : Within ±0.02 mA (for output only)
Error output : With the burnout capability (3.8 mA or 21 mA)
Hold capability : Either the last value hold or the arbitrary value hold is to be selected.
Contact Output
Number of output points : 3
Output type : No-voltage contact output
Contact type : Relay contact, SPDT (1c)
Contact capacity : 250 V AC, 3 A, 30 V DC, 3 A (resistance load)
Contact Capability
R1, R2 : Selectable from upper limit alarm, lower limit alarm, transmission output hold, and cleaning output (opened at alarm operation, closed usually, closed at power-off)
FAIL : Error warning (normally closed; open when an error occurs; open when the power is turned OFF)
Description of Alarm
Setting range : MLSS concentration within the measurable range
Delay time : 0 to 600 s
Cleaning Output
Number of output points : 1
Output type : Active voltage contact output (connected supply voltage output)
Contact type : Relay contact, SPST (1a)
Contact capacity : 250 V AC, 0.5 A
Contact capability : Solenoid valve drive for cleaning
Cleaning Settings
Cleaning frequency : 0.1 h to 168.0 h
Cleaning time : 2 s to 600 s
Hold time : 2 s to 600 s
Timer accuracy : Within 2 min per month
Description of Cleaning Operation
Cleaning operation : Operation of the internal timer
Cleaning operation : Operation of the internal timer and external contact input
Cleaning operation : The internal timer is enabled only at the time of external contact input.
Cleaning trigger : One of the cleaning trigger operations is selected (the internal cleaning sequence starts when external contact input is ON for at least 2 s).
Contact Input
Number of input points : 1
Contact type : No-voltage a contact for open collector
Condition : ON resistance: 100 Ω maximum
Condition : Open-circuit voltage: 24 V DC
Condition : Short-circuit current: 12 mA DC maximum
Contact capability : Can be selected from cleaning directives and transmission hold.
Communication Capability
Type : RS-485
Signal type : Two-wire input/output insulation type (not insulated from transmission output)
Calibration
Calibration method : Zero calibration: With clean water
Calibration method : Span calibration: Concentration conversion method using coefficient input
Working curve selection : Selection of activated sludge and inorganic mud
Adjustment feature : Allows matching the instrument indicated value to a hand analyzed sludge value (Mass method).
Self-check
Sensor check error : Sensor error
Converter error : CPU abnormality, ADC abnormality, memory abnormality
Operating Conditions
Operating temperature range : −20℃ to 55℃ (without freezing)
Operating humidity range : Relative humidity: 5% to 90% (without dew condensation)
Storage temperature : −25℃ to 65℃
Power Source
Rated power supply voltage : 100 V to 240 V AC ±10%, 50/60 Hz
Power consumption : 20 VA (max)
Others
Others : A power switch for maintenance is incorporated.