Hygro-Thermometer Calibration Procedure

2021년 04월 26일 14시 15분 29초

**1\. Range of application** This technical document specifies the procedure for the calibration of hygro-thermometers with humidity measuring ranges from 20

1. Range of application

This technical document specifies the procedure for the calibration of hygro-thermometers with humidity measuring ranges from 20 to 90% RH with 0.1% RH accuracy, temperatures from -20 to 150 ° C, accuracy of 0.1 ° C.

2. Calibrating techniques
- Outside check
- Technical check
- Measurement check
- Measurement Uncertainty Evaluation

3. Calibrating instruments

3.1 Standard measuring instrument (s)

Standard Heating & Cooling Chambers:


- Range of measurement: (-20 ÷ 150) °C

                                         (20 ÷ 95) %RH

- Accuracy: 0,1 °C

                   0,1 %RH

3.2 Other measuring instrument(s)
Hygro-Thermometer:


- Range of measurement: (0 ÷ 50) °C
- Accuracy: ±0.5 °C
- Range of measurement: (10 ÷95) %RH
- Accuracy: ±3%RH

3.3 Auxiliary Instrument(s)
- Water Distillation Unit
- Decimal Stopwatch/Clock
- Alcohol, dedicated gloves, towel

4. Conditions for Calibration
Temperature: (25 ± 2) °C
Humidity: ≤80 %RH

5. Prepare for Calibration
Prior to the calibration, the following preparations must be done:
- Clean the hygrometer to be calibrated.
- Select a standard suitable for the technical requirements of the equipment to be calibrated.
- Turn on the drying of the hygrometer to be calibrated according to the manufacturer's operating instructions.

6. Begin the calibration
6.1 Outside Check
6.2 Technical Check
6.3 Measurement Check

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated Hygro-thermometer is stamped, certified in paper along with calibration results, correction number, and measurement uncertainty at each check point.

b. The recommended calibration cycle of Hygro-thermometer is 12 months.


9. About EMIN's Calibration - Verification - Repair center

EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)

If you want your instruments calibrated, please contact:

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 11, Hoang Cau Moi Street, Trung Liet Ward Dong Da District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


관련 뉴스

How do HRC, HRB, and HRA scales differ in hardness testing?
2026년 06월 15일 14시 30분 32초

HRC, HRB, and HRA all belong to the Rockwell hardness system but are applied to different material groups. Understanding the differences between HRC, HRB, and HRA helps in selecting the correct testing method, accurately reading specifications, and avoiding unnecessary errors in material quality assessment.

Common Errors in Fiber Loss Optic
2026년 06월 13일 09시 24분 06초

Optical power meter used together optical light source, can be applied to determine optical fiber link attenuation — a parameter that reflects the overall quality of the fiber optic cabling system

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결