Dimensions/weight

Maximum weight: 2-module chassis: 6.2 kg (13.6 lb.)

Maximum weight: 4-module chassis: 15.4 kg (33.8 lb.)

Dimensions (L x W x H): 2-module chassis: 46.2 x 19.6 x 25.4 cm (18.2 x 7.7 x 10 in.)

Dimension (L x W x H): 4-module chassis: 47.5 x 43.2 x 27.1 cm (18.7 x 17 x 10.7 in.)

Injectors

Types: Variable volume, variable large volume, backflush, fixed volume

Carrier gas: External cylinder, Helium, hydrogen, nitrogen, argon

GC columns: Wall Coated Open Tubular (WCOT); Porous Layer Open Tubular (PLOT)

Programmable column temperature

Maximum: 250°C or column phase maximum, whichever is lower

Resolution: 0.1°C

Heating rate: 5°C per second maximum, column dependent

Thermal conductivity detector

Linear dynamic range: 10^6±10%

Detection limit: 1 ppm, n-Hexane (WCOT columns)

Internal volume: 240 nL (MEMS)

Repeatability

Retention time: ≤0.1% RSD (WCOT columns)

Peak area: ≤1% RSD (compounds at ≥0.1% concentration, WCOT columns)

Environmental conditions

Operating temperature: 0–50°C ambient

Relative humidity: 5–95% (non-condensing)

Vibration: 2-module chassis: MIL-STD-810F-514.5C, Highway Truck Vibration

Control software: Web-based compatible with common web browsers; Driver for EZ IQ and OpenLAB CDS EZChrom

Communication

Wired Ethernet: RJ-45 connection

Wireless Ethernet: IEEE 802.11a/g/n

Power supply

Power supply input: 100–240 V (ac), 50–60 Hz, 5A

Power supply output: 2-module chassis: 24 V (dc); 10.83A; 260W