Output Current: 3~30A ± (2% ± 2 dgt.)
Test Voltage: AC 6V
Range of Grounding Resistance:
1~300 mΩ, Output Current: 3~10A
1~150 mΩ, Output Current: 10A~30A ± (2% ± 2 dgt.)
Testing Time: 0.1~999.9s ± (2% ± 2 dgt.)
LCD: 16*2 Backlit LCD
Remote Device:
Input: Test, Reset
Output: Pass, Fail, Test-in-Process
Resolving Power:
Current: 0.01A
Resistance: 1mΩ
Test Alarm: Buzzer, indicator light
Correction Method: Software calibration
Memory Device: Memory current, resistance, time, and other settings
Temperature / Humidity: 0℃~40℃ ≤ 75%RH
Standard Accessories: JK26003A
Dimensions/Weight: W×H×D: 280mm × 100mm × 380mm / 10kg
Manual
JK7305 ground resistance tester is used to measure the internal electrical equipment grounding resistance, it reflects the total grounding of electrical equipment around the exposed conducting part and electrical equipment terminal (contact resistance). Ground resistance tester in order to eliminate the effects of contact resistance test, using four terminal measurement method, namely in the measured electrical exposed between the guide part and the main earthing terminal with current voltage measurement, and then it ends, calculate the resistance value.
l Constant current linear amplifier output
l The front panel correction, high accuracy
l Standard RS232, standard PLC interface
l Key lock function
l Four - end measurement precision
l Up to 15 sets of test pattern memory
l Alarm function