인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

Flow Rate: 28.3L/min ±2%
Screen: 7-inch color touch screen
Particle Size: 0.1, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0, 5.0μm
Counting Efficiency: 0.1μm:50%±20%;>0.15μm:100%±10%
Error of Particle Size: 0.5μm、5μm ≤±30%
Concentration Error: 0.5μm≤±30%FS
Repeatability: ≤10%FS at the same test condition
Laser Source: Solid-state laser
Zero Count: <1 count /5min
Sample Mode: Manual, automatic; accumulation counting, partition counting; Concentration mode, number mode; UCL calculation
Alarm: Sound and visual alarm, low power, abnormal or other situations happen.
Data Storage: 8GB,about 100000 groups
Area/room/location/recipe: Area, room, location, recipes can be freely combined
Sampling Delay: 0~1000min
Sampling Time: 1s~1000min
Sampling Volume: 0.47L~28300L
Sampling Frequency: 1~1000
Sampling Interval: 1~1000min\
Operation Environment: (10~40)°C /,(20~95) %RH no condensation
Storage Environment:(-20~50)℃, ≤85%RH
Date: automatic recording; USB export; build in printer
Power Supply: AC(100 ~240)V,(50~60) Hz
Battery: continuous sampling ≥4 h
Noise:<60dB(A)
Reference Standards
GMP Good manufacturing practice of medical products
ISO 14644 Cleanrooms and associated controlled environments
ISO 21501 Light scattering airborne particle counter for clean spaces
JJF 1190-2008 Calibration specification for airborne particle counters
JIS B 9921:2010 Light scattering airborne particle counter for clean spaces
GB/T 16292-2010 Test method for airborne particles in clean room (zone) of the pharmaceutical industry
GB/T 6167-2007 Methods of testing the performance of airborne particle counter
Features
Up to 9 test channels, accumulation counting, partition counting, concentration mode, number mode
7-inch touch color screen, friendly human-computer interaction
Preset different cleanroom standards, can automatically judge whether qualified
Output test report, support UCL calculation
Three-level user management and audit tracing to ensure data integrity
The preset recipes can be assigned to rooms in different areas
Built-in high-precision fan, stable flow control
Built-in HEPA filter to filter the exhaust
Automatic zero count without plugging in external HEPA filter
Support exporting data to USB flash drive and printing data by built-in printer; equipped with a rich communication interface
Ambient temperature, humidity and atmospheric pressure parameters can be recorded during test.