Applications
Medical facilities and cleanrooms
Biosafety cabinets and fume hoods
Independent filter certifiers
Pharmaceutical manufacturers
Specification
Sampling flow: 28.3L/min,±2.5%
Concentration detection range: 0.01~125μg/L; Ultra-low concentration can be extended to 0.0001 μg/L; Ultra-high concentration can be extended up to 700 μg/L
Detection of leakage rate: 0.0001%~100%
Detection precision: 1% of the value in the range of 0.01% to 100%
Detection repeatability: 0.5% of the value in the range of 0.01% to 100%
Data storage capability: 100000 groups
Size(W×D×H): (300×330×184)mm
Weight: 8.9 kg
Packed weight: 15kg(Outer packing, instrument, accessories, etc.)
Power: AC110V/60Hz
Power consumption: <150W
Environmental temperature: 10℃~35℃
Environmental humidity: 5%~85%RH(No dew, no ice)
Storage requirements: -10℃~ 40℃, RH<85%, no dew
Test medium: PAO, DOP and other aerosol types
Noise:<65dB(A)
Standards
NSF/ANSI 49-2019: Biosafety cabinetry
ISO14644-3:2005: Cleanrooms and associated controlled environments—Part 3:Test methods
GB 50073-2013: Code for design of clean workshop
GB 50591-2010: Code for construction and acceptance of cleanroom
2010 GMP for pharmaceutical products-plant and equipment
YY0569-2005: Biosafety cabinet
JJF 1800-2020: Calibration specification for aerosol photometer
JJF 1815-2020: Calibration specification for class II biosafety cabinet
Features
• Powerful function
Stable digital photometer.
Dynamic Range: 0.0001μg/L~700μg/L.
Data query
USB & Printer available for real-time reporting.
Audit Tracing & assign permissions, further guarantee data integrity.
Sampling data can be imported to PC.
• Automatic reminder
When exceed the set value, light & voice alarm.
Self protection in case of failure.
• Good human-computer interaction
5.0 -inch color screen, touch operation.
Lightweight, portable and equipped with suitcase, easy to carry.
Built-in battery(optional)≥ 3.5H.