Minimum Size Detection: 0.5 μm
Standard channel sizes: 0.5, 5.0 μm
Optional channel sizes: 0.5, 1.0, 5.0, 10.0 μm
Additional Available sizes: 0.7, 3.0 μm
Flow Rate: 1 CFM (28.3 LPM), 1/4” fittings
Weight: 10.9 lbs (4.76 kg)
Concentration Limits: 1,000,000 Particles/ft3@10% coincidence loss Suitable for ISO classes 1-8 cleanrooms
Communication: Ethernet or Serial.
Modbus Protocols: TCP or RTU or ASCII
Location Labels: 200 Location Labels
Data Storage Records: 3,000
Reports: FED 209E, ISO 14644-1:1999, ISO 14644-1:2015, & EU GMP regulatory pass/fail reports on screen or printer
Secure Data Transfer: Designed for 21 CFR part 11 Compliance
Languages: English, Spanish, German, French, Italian, Russian, Simplified Chinese, Japanese, Korean
Self Diagnostics: Laser power supply, laser current, laser power, photo detector power supply, background voltage, photo detector health.
Web Server: Remote access via web browser
Ports: Ethernet, RS485 (RJ-45), USB
Battery: 4 hours Nominal, 3 hours continuous use
Enclosure: 304 Stainless Steel
Power: 24 VDC, 5 A Max Draw
Vacuum Source: PID Controlled Vacuum Pump
Air Sample Output: Internally Filtered To HEPA Standard (>99.97% @ 0.3μm)
Sample Count Modes: Automatic, Manual, Beep, Concentration
Light Source: Extreme Life Laser Diode
Calibration: ISO 21501-4 Compliant and Accredited ISO 17025
Zero Count Level: < 1 Count / 5 minutes (meets specifications of JIS B 9921 and reporting requirements of ISO 21501-4 Annex C)
Dimensions: 9.00”x 6.69“ x 8.46” (22.86 x 16.99 x21.48 cm)
Software: LMS Pro or LMS Xchange
Included: Power supply 24 VDC, power cord, iso kinetic probe, Operating Manual on USB key