• The LS9956 can test for Withstand Voltage (AC/DC), Insulation Resistance (IR), Leakage Current (LLC), Ground Resistance (GR) and Power

Withstand Voltage Test (AC/DC)
Voltage Range: 100~4000V
Accuracy: ± (5%+3V)
Current Range:
AC: 0.10~20.00mA
DC: 0.10~12.00mA
Leakage Current Test (LLC)
LLC Range: 0.10~20mA
Accuracy: ± (0.3%+5uA) 
Voltage Range: AC10.0~300.0V
Insulation Resistance Test (IR)
IR Range: 1.00~2000MΩ
Accuracy: 5% + 0.5MΩ
Voltage Range: DC100~1000V
Grounding Resistance Test (GR)
GR Range: 0~600mΩ
Accuracy: ± (5%+2mΩ)
Output Current: AC1.00~30.00A

Power Test
Voltage: 10~300V
Current: 0.010~20.00A
Power: 1.0~6000.0W
PF: 0.2~1.0
Accuracy: Class 0.5
Test Time: 0~999.9s

Datasheet



• Remote control device (Optional)

• Display all setting parameters and testing results in the same big LCD menu

• PASS/FAIL limit values can be set. It has alarm sound and indicator light

• Test mode programmable; 50 groups setting available, 8 test steps per group

• Quick discharge

• Support 50Hz and 60Hz frequency; Electric safety wall detecting feature

• The LS9955 can do test for Withstand Voltage (AC/DC), Insulation Resistance (IR), Leakage Current (LLC) and Ground Resistance (GR)

• The LS9956 can test for Withstand Voltage (AC/DC), Insulation Resistance (IR), Leakage Current (LLC), Ground Resistance (GR) and Power

 