Compatible Detectors: Up to four external GM, scintillation, proportional, or neutron detectors

Internal Detector / Range: N/A

High Voltage: 300 to 2400 Vdc

Linearity: Readings within 10% of true value

Modes: Four operational modes including rate, max, integrated dose, and count

Display: 3-digit auto-ranging LCD with large 20 mm (0.8 in.) digits.

Units: (k)(M)c, (k)cps, (k)(M)cpm, (k)(M)d, (k)(M)Bq, (k)(M)Bq/cm2, (k)(M)dpm, (µ)(m)(k)(M)R, (µ)(m)(k)R/h, (µ)(m)(k)(M)Sv, (µ)(m)Sv/h, (µ)(m)(k)(M)Rem, (µ)(m)(k)Rem/h.

Indicators: bar graph, radiation type (alpha (α), beta (β), gamma (γ), neutron (n)), mode (MAX, INTG), three alarm levels (ALARM 1,2,3), fault messages (FAULT, OVERLOAD, LOSS OF COUNT, OVERRANGE), USB, wireless, audio, low-battery.

Display Range: 

00.0 cps to 999 kcps

00.0 cpm to 99.9 Mcpm

00.0 Bq to 999 MBq

00.0 Bq/cm2 to 999 MBq/cm2

00.0 dpm to 999 Mdpm

0.00 µR to 999 MR

0.00 µR/h to 9.99 kR/h

0.00 µSv to 999 MSv

0.00 µSv/h to 999 Sv/h

0.00 µRem to 999 MRem

0.00 µRem/h to 9.99 kRem/h

0 c to 999 Mc (counts)

0 d to 999 Md (disintegrations)

Operational display ranges are dependent on the detector and dead time configuration. The minimum and maximum display can be set to limit the display to the calibrated range.

Backlight: Built-in ambient light sensor automatically activates LED backlight unless internal dipswitch is set to continuous-on

User Controls: 

ON/AUD: Extra-long press to turn ON; Short press to adjust audio level; Extra-long press plus 3 seconds to turn OFF

UNITS: Short press Changes the units between count rate, dose/exposure, or activity

DETECTOR: Short press to change active external detector to the next enabled detector. On 3003i, extra-long press to switch between internal and external detector.

MODE: Short press alternates between modes: RATE (displays count rate), MAX (captures peak rate), COUNT (user-selectable preset count time from 0 to 31.68 years), and DOSE (current accumulated dose)

ACK/RST: Short press to acknowledge alarms, extra-long press to reset current mode

LOG: Short press to log current display (requires Lumic Data Logger Kit) or initiate count in COUNT mode, extra-long press to change count in COUNT mode

PRG: Short press in first 4 seconds of power-up will grant access to setup menu

Response Time: User-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), fixed FAST (4 s)

Alarms: Three count rate, exposure/dose, scaler, and accumulated dose alarm set-points adjustable over the display range

Self-Diagnostics: Detects and reports internal instrument issues, indicated by graphic FAULT display and audio warning

Overload Protection: High count rate saturation protection prevents false display of lower count rates. OVERLOAD will appear on the display and an audio warning will be triggered.

Loss of Count Protection: After a user-defined time, typically 60 seconds, with no pulses from detector, the unit will flash a zero reading, an audio alarm will be triggered, and LOSS OF COUNTS will appear on the display

Overrange: If reading should exceed the predefined detector range, the unit will flash a maximum reading, an audio warning will be triggered, and OVERRANGE will appear on  the display

Audio: "Click" audio greater than 65 dB at 0.6 m (2 ft.), Alarm audio greater than 72 dB at 0.6 m (2 ft.), approximately 4.5 kHz

Calibration Protection: Internal dipswitch to disable front-panel calibration parameter changes

Power: Four "AA" batteries

Environmental Rating: NEMA rating of 3x or IP rating of 53

Size (H x W x L): 16.5 x 11.4 x 21.6 cm (6.5 x 4.5 x 8.5 in.)

Weight: 0.9 kg (2.0 lb) (excluding detectors and clips)