인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

DC voltage: 0 - 1000 V
± 10 ppm / year
AC voltage: 0 - 1000 V, 20 Hz - 100 kHz
sinusoidal & non-sinusoidal waveforms
± 180 ppm / year
DC current: 0 - 30 A (1500 A with 140-50 Current Coil)
± 100 ppm / year
AC current: 1 uA - 30 A, 20 Hz - 10 kHz (1500 A with 140-50 Current Coil)
sinusoidal & non-sinusoidal waveforms
± 500 ppm / year
DC Power & Energy: 0.4 mW - 4.8 kW (240 kW with 140-50 Current Coil)
1-1999 s
± 510 ppm / year
AC Power & Energy: 0.4 mW - 4.8 kW, 40 - 400 Hz (240 kW with 140-50 Current Coil)
1-1999 s, power factor -1 to +1
± 635 ppm / year
Resistance: 0 Ω - 1 GΩ, resolution 4½ digits
± 100 ppm / year
Capacitance: 700 pF - 100 uF, resolution 4½ digits
± 0.5 % / year
Temperature: TC BEJKNRST with cold junction compensation
RTD ± 0.07°C / year
Frequency: 0.1 Hz - 20 MHz
± 50 ppm / year
Process Multimeter
DC voltage: 0 - 20 V
mVDC resolution down to 100 nV
± 100 ppm / year
DC current: 0 - 25 mA
± 150 ppm / year
Resistance: 0 Ω - 2.5 kΩ, resolution 4½ digits
± 200 ppm / year
Temperature: TC BEJKNRST
RTD ± 0.1°C / year
Frequency: 1 Hz - 15 kHz
± 50 ppm / year
Accessories: Adapter 140-01, Current Coil 140-50, Adapter 40, Adapter 60