Temperature
Setting temperature range: +20 to +80 °C
Working temperature range: min. 10°C above ambient up to +80°C
Setting accuracy temperature: 0.1 °C
Temperature sensor: 1 Pt100 sensor DIN class A in 4-wire-circuit
Control technology
ControlCOCKPIT: SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable temperature values
adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
Adjustable parameters: temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime
Ventilation
Fan: forced air circulation in working chamber, fixed value 100 %
Stainless steel interior
Dimensions: w(A) x h(B) x d(C): 1040 x 720 x 600 mm (d less 39 mm for fan)
Interior: easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume: 449 l
Max. number of internals: 8
Max. loading of chamber: 300 kg
Max. loading per internal: 30 kg
Electrical data
Voltage Electrical load: 230 V, 50/60 Hz approx. 1800 W
Ambient conditions
Set Up: The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation: max. 2,000 m above sea level
Ambient temperature: +5 °C to +40 °C
Humidity rh: max. 80 %, non-condensing