Metrel CS 2890 Power Calibrator/Simulator (350 Vrms; 2000 A)
TECHNICAL SPECIFICATION - CALIBRATOR
Warmup time: Minimum 30 minutes, connection to external power supply required
Settling time: Less than 10 seconds
Reference temperature: 23 °C ± 2 °C
Voltage/current stability: ±0.06 % (90 days)
Calibration currents: 0.05 – 0.1 – 0.2 – 1 – 2 V/ ±0.1 %
Calibration voltages: 5-11-14-23-50-75-110-150-165-206-230-250-345-400-500 V/ ±0.1 %
Voltage selection: 5-11-14-23-50-75-110-150-165-206-230-250-345-400-500 V
Frequency: 50 Hz
Uncertainty / 90 days: ±0.06 %
Setup resolution (under Adjustment menu): 0.0001 V
TECHNICAL SPECIFICATION - SIMULATOR
Fundamental RMS voltage output
Output voltage AC: 50 ... 350 V
Resolution: 10 V
Accuracy: ±0.1 %
Event RMS voltage output
Output voltage AC: 0 ... 350 V
Resolution: 0.01 V
Accuracy: ±2 %
Fundamental RMS current
Range: A 1033 (1 A ... 2000 A)
Output voltage: 1 mV ... 1 V
Overall current accuracy: ±0.1 %
Inrush RMS current output
Inrush current
Range 1: 2.0 mVRMS ... 200.0 mVRMS, Accuracy: ±0.5 % · URMS, Crest factor: 1.5
Range 2: 20.0 mVRMS ... 2.0000 VRMS, Accuracy: ±0.5 % · URMS, Crest factor: 1.5
Frequency
Output range: 45 Hz ... 70 Hz
Resolution: 1 Hz
Accuracy: ±10 mHz
General
Measuring category CAT I / 300 V
Dimensions: 230 x 140 x 80 mm
Weight (with batteries): 1.36 kg
Display Colour: 4.3 TFT liquid crystal display (LCD) with backlight, 480 x 272 dots.
Batteries: 6 x 1.2 V NiMH rechargeable batteries type HR 6 (AA)
Working temperature range: 0 °C ... +40 °
*For detailed specifications, please refer to the datasheet.