For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Michell Instruments EA2-OL-AAAAA Online Dew-Point Hygrometer (-100 ~ +20°C)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Measurement Range: -100 to +20°C (-148 to +68°F) dew point

Accuracy: ±2°C (±3.6°F) dew point*

Response Time: 5 mins to T95 (dry to wet)

Repeatability: 0.5°C (0.9°F) dew point

Sensor Calibration: Traceable 13-point calibration certificate


Electrical Specifications

Online Output Signals: 4–20 mA

Online Output: Dew point or moisture content

Maximum Analog Output Scaled Range: Dew point: -110 up to +20°C (-166 up to +68°F); Moisture content in gas: 0 up to 3000 ppmV

Online Supply Voltage: AC: 85 to 264 V

Online Current Consumption: 60 mA maximum

Electrical Safety: EN61010-1


Operating Specifications

Operating Temperature: Sensor: -40 to +60°C (-40 to +140°F); Monitor: 0 to +50°C (+32 to +122°F)

Compensated Temperature Range: Sensor: -20 to +50°C (-4 to +122°F); Monitor: not applicable

Storage Temperature: Sensor: -40 to +60°C (-40 to +140°F); Monitor: -10 to +60°C (+14 to +140°F)

Operating Pressure: 52.5 MPa (525 barg / 7614 psig) maximum

Sensor Flow Rate: 1 to 5 Nl/min mounted in standard sampling block; 0 to 10 m/sec direct insertion


Mechanical Specifications

Ingress Protection: Sensor: IP66 in accordance with BS EN 60529:1992+A2:2013; NEMA 4 protection in accordance with NEMA 250–2014; Monitor: IP65 (NEMA 4X) front panel only

Sensor Oxygen Service: Optional: Cleaned for enriched oxygen service

Sensor Housing Material: 316 stainless steel

Dimensions: Sensor: L=132 mm × Ø45 mm / L=5.2” × Ø1.77” (with connector cable); Monitor: 1/8 DIN Case, 96 × 48 × 85 mm / 3.78 × 1.89 × 3.35” (W × H × D)

Filter (Sensor Protection): Standard: HDPE <10 µm; Optional: 316 stainless steel sintered guard <80 µm

Sensor Process Connection: 5/8” - 18 UNF; KF40 Flange

Accessories: Flange for 5/8” - 18 UNF process connections

Sensor Weight: 150 g (5.29 oz)

Sensor Electrical Connections: MiniDIN 43650 Form C

Online Sensor Cable: 0.8m connector/cable


Sensor Diagnostic Conditions (factory programmed)

Sensor Fault: 23 mA

Under-range Dew Point: 4 mA

Over-range Dew Point: 20 mA