인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

Size Range: 0.3 to 25 μm
Size Channels: Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning
Flow rates: 1.2 LPM
Concentration Limits: 27,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 50,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Aerosol Concentration Range: 0.01 to 20,000 μg/m³
Light Source: Long life laser diode
Counting Efficiency: 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction. No appreciable drift.
Count Modes: Particles/second, Real-Time Meter, cumulative/diff. (count/m³ and count/ft³) and mass concentration (PM)
Count Alarms: 1 to 9,999,999 counts
Calibration: NIST traceable
Display: 4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Optional Printer : External thermal printer available
Vacuum Source: Internal pump with automatic flow control
Filtered Exhaust: Internal HEPA filter
Number of Channels: 6
Custom Size Channels: Calibration for custom size channels available
Audible Alarm: Adjustable integrated alarm
Reports: ISO 14644-1, EU GMP Annex 1, FS 209E
Recipes: 50 user-configurable recipes
Communication Modes: Ethernet, USB, Wireless 802.11 b/g (Optional), RS485 or RS232
Optional Environmental Sensor: Temperature 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC) and relative humidity probe 15-90% ±2%
Alarm: Alarms on counts for all particle sizes, sensor failure, environmental sensors and flow rate
Standards: ISO 21501-4 and JIS B9921
Calibration: Recommended minimum once per year
External Surface: Stainless steel
Dimensions (L x W x H): 5.22 x 2.25 x 9.13 in (13.3 x 5.7 x 23.2 cm) includes barb fittings
Weight: 2.6 lb (1.18 kg)
Accessories: User manual + Instrument Management Software (IMS) on USB flash drive, isokinetic probe, zero-count filter, USB cable, and power supply
Optional Accessories: Printed manual, external printer and isokinetic probes
Buffer Memory: Securely stores up to 45,000 records with particle counts, environmental data, locations, and timestamps
Sample Locations: Up to 1,000 locations 20 characters long
Sample Time: 1 second to 99 hours
Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions: 41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions: 32-122ºF (0-50ºC) / Up to 98% non-condensing