Size Range: 0.3 to 25 μm
Size Channels: Factory calibrated at 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm variable binning
Flow rates: 1.2 LPM
Concentration Limits: 27,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 50,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Battery Run Time: >10 hours continuous operation
Light Source: Long life laser diode
Counting Efficiency: 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction.
Count Modes: Real-Time Meter and graph, cumulative/differential count/m³and count/ft³, and mass concentration (PM)
Count Alarms: 1 to 9,999,999 counts
Calibration: NIST traceable
Display: 4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Printer (Optional): External thermal printer available
Vacuum Source: Internal pump with automatic flow control
Filtered Exhaust: Internal HEPA filter
Number of Channels: 6
Custom Size Channels: Calibration for custom size channels available
Audible Alarm: Adjustable built-in alarm
Battery: Removable Li-ion
Battery Recharge Time: <2 hours
Reports: ISO 14644-1, EU GMP Annex 1, FS 209E
Recipes: 50 user-configurable recipes
Communication Modes: Ethernet, USB, Wireless 802.11 b/g (Optional), RS485 or RS232
Environmental Sensor: Includes temperature (32-122°F, ±1°F / 0-50°C, ±0.5°C) and relative humidity probe (15–90% RH, ±2%)
Alarm: Alarms for particle counts (any size channel), sensor failure, environmental sensors, and flow rate
Standards: ISO 21501-4 and JIS B9921
Calibration: Recommended minimum once per year
External Surface: Stainless steel
Dimensions (L x W x H): 5.22 x 3.81 x 9.42 in (14.0 x 9.7 x 23.9 cm) includes barb fittings
Weight: 4 lb (1.8 kg)
Accessories: User manual + Instrument Management Software (IMS) on USB flash drive, isokinetic probe, temperature relative humidity sensor, zero-count filter, battery, USB cable, and power supply
Optional Accessories: Printed manual, spare battery, external battery charger, external printer and isokinetic probes
Buffer Memory: Securely stores up to 45,000 records with particle counts, environmental data, locations, and timestamps
Sample Locations: Up to 1,000 locations 20 characters long
Sample Time: 1 second to 99 hours
Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions: 41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions: 32-122ºF (0-50ºC) / Up to 98% non-condensing