Accuracy: Cx: ±(reading*1%+1pF), tgδ:
±(reading*1%+0.0040)
Anti-interference index: frequency conversion
anti-interference, can still reach above precision at 200% interference
Capacitance range:
3pF~60000pF/10kV 60pF~1μF/0.5kV (test with
internal voltage)
3pF~1.5μF/10kV 60pF~30μF/0.5kV (test with external
voltage)
Resolution: 0.001pF
Tgδ range: no limit, resolution: 0.001%, capacitance,
inductance and resistance auto identification
Test current range: 10μA~1A
Test with internal voltage:
Voltage setting range: 0.5~10KV
Max current output: 200mA
Test frequency: 45Hz, 50Hz, 55Hz, 60Hz, 65Hz (single
frequency)
45/55Hz,
55/65Hz, 47.5/52.5Hz (Dual frequency)
Frequency resolution: ±0.01Hz
Test with external voltage
Max test current : 1A (measurement of ungrounded test
object), power frequency or conversion frequency: 40~70Hz
Max test current: 10KC/1A (measurement of grounded test
object), power frequency or conversion frequency: 40~70Hz
CVT self-excitation low-voltage output
Output voltage: 3~50V, output current: 3~30A
Test time: about 40s (related with test mode)
Input power: 180V~270V (AC), 50Hz±1% (supply by city power
or electric generator)
Printer: A7 micro-thermal printer
PC port: standard RS232 port
Working conditions: -10℃~50℃ (temperature), <90% (R.H.)
Dimension (mm): 460*360*350
Weight: 28KG
Dielectric loss test is a effect way to detect the damp degeneration and local defect of electrical equipment insulation, it is used in transformer winding and bushing, current transformer, voltage transformer, breaker, high-tension switch, parallel capacitor, coupling capacitor and etc. Dielectric loss detector adopts frequency conversion power technology, and make use of single chip microcomputer and electronic technique to carry out automatic frequency conversion and data operation, to make the detector has the advantage of strong anti-interference, fast test speed, high accuracy, fully automatic digitization and simplification operation. Take the high power switch power supply to output 45Hz and 55Hz pure sine-wave, and automatic inflates, can supply 10KV voltage at most. Automatically filter 50Hz interference, it is suitable for site test of transformer substation.