pH Tester Calibration Procedure

2021년 04월 26일 13시 50분 20초

**1\. Range of application** This technical document specifies the calibration of pH tester with glass electrodes of (0-14) pH in measuring range and accurac

1. Range of application

This technical document specifies the calibration of pH tester with glass electrodes of (0-14) pH in measuring range and accuracy of at least 0,02 pH

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

Technical Calibration Buffer:


- Range of measurement: (0-14) pH

- Measurement Uncertainty: ≤ 0.01 pH

3.2 Other measuring instrument (s)

Heated Digital Water Bath:


- Range of measurement: (+5 ÷ 99) °C

Standard Humidity, Temperature Meter


- Range of measurement: (-20 ÷ 70) °C

- Accuracy:

  ± 0.2 °C (10 ÷ 40 °C)

  ± 0.4 °C of the remaining

  ± 1%RH (10 ÷ 90) %RH 

  ± 2.0 %RH (of the remaining)

Hygro-Thermometer

- Range of measurement: (0 ÷ 50) °C

- Accuracy: ±0.5 °C

- Range of measurement: (10 ÷95) %RH

- Accuracy: ±3%RH

3.3 Auxiliary Instrument(s)

- Distilled water

- Sprayer

- Tissue, glass cup

4. Conditions for Calibration

Temperature: (25 ± 2) °C

Humidity:  80 %RH

The temperature, as well as the humidity, must be recorded before and after calibration

5. Prepare for Calibration
Prior to the calibration, the following preparations must be done:
- Prepare the Technical Calibration Buffer
- Place the pH tester and Technical Calibration Buffer in the calibration room at least 1 hour prior to calibration. 
- The Technical Calibration Buffer must be kept at constant temperature at least 1 hr at (25 ± 0,2) 0C in Heated Digital Water Bath
- Prior to calibration, immerse the pH tester electrode with distilled water at least 3 hrs in the lab.
Before measuring the pH level of Technical Calibration Buffer, the followings must be met:
- Clean the pH tester’s electrodes with distilled water.
- Dry the electrode with tissue.
The pH tester to be calibrated must be operating at least 15 mins prior to calibration.

6. Begin the calibration
6.1 Outside Check
6.2 Technical Check
6.3 Measurement Check
A pH tester is checked and measured according to the following content, method and requirements:
- Calibrate the pH tester according to the manufacturer's technical documents.
- Check the error with standard pH solutions
- Measure at least 3 times for each pH standard solution. Record the measured results in the record.
- Calculate the error

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated pH tester is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of pH tester is 12 months.

9. About EMIN's Calibration - Verification - Repair center

EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)

If you want your instruments calibrated, please contact:

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220

Further information about pH tester calibration service can be found here.


관련 뉴스

Can an analog multimeter still measure capacitors?
2026년 08월 05일 15시 30분 41초

An analog multimeter can still be used to quickly check if a capacitor is shorted, open-circuited, or capable of charging and discharging. However, if precise capacitance measurement or an assessment of capacitor quality is required, a digital multimeter or LCR meter will yield more reliable results

Measuring Pump Shaft Speed? When Is It Necessary, and How Should It Be Done Correctly?
2026년 08월 04일 14시 45분 15초

Measuring pump shaft speed helps determine the equipment's actual rotational speed (RPM), allowing for comparison against design specifications to detect issues such as belt slippage, incorrect gear ratios, abnormal motor operation, or failure to achieve the desired flow rate. Depending on operating conditions, engineers may choose a laser tachometer, a contact tachometer, or a stroboscope

PONOVO Conducts Product Training at EMIN Vietnam
2026년 08월 03일 11시 24분 20초

To share knowledge about testing equipment and testing solutions in the electrical and energy sector, Ponovo organized a training and product presentation session at the EMIN office on July 9, 2026.

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결