For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0

Digital MultiMeter Calibration Procedure

2021년 04월 06일 10시 26분 04초

**1.Range of application** \- DC Voltage 1000V \- AC Voltage 1000V, Frequency 100kHz \- DC Current 30A \- AC Current 30A, Frequency 10kHz \- Resistor 100

1.Range of application

- DC Voltage 1000V

- AC Voltage 1000V, Frequency 100kHz

- DC Current 30A

- AC Current 30A, Frequency 10kHz

- Resistor 1000MΩ

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s): Multifunction Calibrator (MFC)

Meatest M142 Multifunction Calibrator 15 ppm

- DC Voltage: 0µV - 1000V.

- AC Voltage: 0,1mV - 1000V, Frequency: 20Hz - 100kHz.

- DC Current: 0µA - 30A.

- AC Current: 1µA - 30A, Frequency: 20Hz - 10kHz.

- Resistor: 0 Ω - 1GΩ.

3.2 Auxiliary Instrument(s)

3.2.1 Testo Hygro-Thermometer 608-H1

Testo 608-H1 - Thermal Hygrometer (Part Number 0560 6081) | TEquipment

- Range of Measurement: (0 ÷ 50) °C; Accuracy: ±0,5 °C

- Range of Measurement: (10 ÷ 95) %RH; Accuracy: ±3 %RH

3.2.2 Dedicated Measuring Probe

- Probes with small electromotive force.

4. Conditions for Calibration

- Temperature: (23 ± 5) °C.

- Humidity: no higher than 80% RH.

5. Prepare for Calibration

- Make sure the power supply to the DMM meets the requirements specified in the catalog, the fuses, the protective circuit are still working properly.

- The DMM should be placed in the calibration environment for at least 8 hours.

- MFC and DMM must be turned on and dried for at least 30 minutes prior to calibration.

6. Begin the calibration

a. Outside Check

b. Technical Check

c. Measurement Check

    1. General requirements

    2. Begin the DMM Calibration

        a. DC Voltage

        b. AC Voltage

        c. DC Current

        d. AC Current

        e. Resistor

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated DMM is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of DMM is 12 months.


For calibration of DMM, click here.

For further information about Digital MCalibration service, please contact

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


관련 뉴스

What Is Resonant Frequency? Why Can a Single Frequency Range Damage the Entire Product?
2026년 07월 10일 15시 10분 46초

A device may operate stably for months, yet suffer cracks, broken solder joints, or snapped component leads after being transported. What surprises many engineers is that the vibration amplitude during transport is no greater than usual. The root cause lies in a specific frequency range where the product's structure absorbs energy most intensely, triggering resonance and causing continuous stress concentration at a single point.

Should Carton Packaging Undergo Vibration Testing or Drop Testing First?
2026년 07월 10일 10시 38분 40초

Một kiện hàng gần như không bao giờ bị rơi ngay khi vừa rời khỏi nhà máy. Quãng đường hàng trăm kilomet trên xe tải, container hay băng chuyền mới là giai đoạn diễn ra đầu tiên, nơi bao bì liên tục chịu rung lắc trong nhiều giờ. Chỉ sau đó mới đến những cú va chạm khi bốc xếp, giao nhận hoặc rơi ngoài ý muốn.

Distinguishing Between DFT and WFT in Coating Thickness Measurement
2026년 07월 09일 14시 40분 30초

Many projects encounter a situation where a freshly applied coating looks perfectly uniform and paint consumption matches the estimate, yet additional coating is required during inspection because the thickness is insufficient. This leads many to suspect measurement errors or equipment inaccuracy, whereas the actual cause is a confusion between WFT (Wet Film Thickness) and DFT (Dry Film Thickness)

Can a car paint thickness gauge detect if a vehicle has been in a collision?
2026년 07월 08일 16시 56분 01초

Many used car buyers often spend a considerable amount of time examining the paintwork, the gaps between body panels, or the overall condition of the exterior. However, these signs can be skillfully masked following repairs; precise color matching and careful polishing can make the vehicle appear almost factory-original.

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결