Metrological parameters
-Maximum capacity [Max]: 310 g
-Minimum load: 10 mg
-Readability [d]: 0.1mg
-Verification unit [e]: 1 mg
-Tare range: -310 g
-Standard repeatability [5% Max]:0.05 mg
-Standard repeatability [Max]: 0.1 mg
-Standard minimum weight (USP): 100 mg
-Standard minimum weight (U=1%, k=2): 10 mg
-Permissible repeatability [5% Max]: 0.07 mg
-Permissible repeatability [Max]: 0.15 mg
-Linearity: ±0.3 mg
-Eccentric load deviation: 0.3 mg
-Sensitivity time drift: 1×10⁻⁶/Year×Rt
-Stabilization time: 1,3s
-Adjustment: internal (automatic)
-OIML Class: I
Physical parameters
-Leveling system: automatic - Reflex Level System
-Display: 10″ graphic colour touchscreen
- Weighing chamber: automatic
- Weighing chamber doors: automatic
- Delivery components: Analytical Balance, weighing pan, weighing pan shield, centring ring, brush, fabric dust cover, power supply.
-Weighing pan dimensions: ø100 mm
- Weighing chamber dimensions: 200×170×220 mm
- Net weight: 14.7 kg
- Gross weight: 19.1 kg
- Protection class: IP 43
Communication interface
-Communication interface: 2×USB-A, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot
-Electrical parameters: Power supply
- Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A
- Balance: 12 – 15V DC 1.6A max*
Environmental conditions
-Operating temperature: +10 — +40 °C
-Operating temperature change rate: ±0.3°C/1h (±1°C/8h)
- Relative humidity: 40% — 80%
- Relative humidity change rate: ±1%/h (±4%/8h)