Test device: Automatic range conversion, true RMS measurement

Medical GB9706.1-2020 Standard

Leakage current measurement range AC: I (3~99.9)uA Resolution: 0.1uA

Ⅱ (100.0~999.9)uA Resolution: 0.1uA

Ⅲ   (1000~9999)uA  Resolution:1uA

Patient current, patient auxiliary current: DC Measurement range: (3~999.9)uA Resolution: 0.1uA

Measurement accuracy: ±5%+3 digits Note: The accuracy range is current > 10uA or more

Frequency response range: DC ~ 1MHz Measurement impedance circuit (MD): in line with GB9706.1-2020

Laboratory GB4793.1-2007 Standard

Leakage current measurement range AC: I (3~99.9)uA Resolution: 0.1uA

Ⅱ  (100.0~999.9)uA  Resolution:0.1uA

Ⅲ   (1000~9999)uA  Resolution:1uA

Measurement accuracy: ±5%+3 digits Note: The accuracy range is current > 50uA or more

Measuring Impedance Circuit (MD): Compliant with GB4793.1-2007

The output voltage

Measure the voltage output range of the power supply: 0V ~ 300V, the resolution is 1V, the accuracy: ± (5% + 2 words)

Output voltage capacity: 5000VA

Current upper limit setting

Range: (3~9999)uA 

Resolution: 1uA Accuracy: ±(4%+3 words) 

The accuracy range is current>10uA

Timing device

Range: (3~9999)S 

Resolution: 1S 

Accuracy: ±5%

Operating temperature: 0-40℃  ≤75%RH

Power requirements: 220±10%  50Hz/60Hz±3Hz

Interface: Standard RS232, PLC

Memory group: five sets of memory

Screen: LCD2004C

Form factor(D×H×W): 430mm×150mm×350mm

weight: 35KG

Instrument power consumption: <50W

Random standard accessories: power cable RK00001、RS232 Communication Cable RK00002、RS232 to USB cable RK00003、16G U disk (instruction manual)、Cable interface transfer driver CD、test leadRK00049/RK00050