인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

Conductivity:
Range: 0.000 μS/cm to 1000 mS/cm
Resolution: 0.001 μS/cm minimum, various with range selection
Accuracy: ±1.0% FS
Reference Temperature: 20, 25 ºC
Calibration Points: Up to 3
Standard Recognition: 84μS/cm, 1413μS/cm, 12.88mS/cm
Resistivity:
Range: 5.00 Ω·cm~20.00 MΩ·cm
Resolution: 0.01 Ω·cm minimum
Accuracy: ±1.0% FS
TDS:
Range: 0.00 ppm ~300 ppt
Resolution: 0.01ppm minimum, various with range selection
Accuracy: ±1.0% FS
Salinity:
Range: (0.00~8.00)%
Resolution: 0.0001
Accuracy: ±0.2%
Temperature:
Range: -5.0 to 110 ºC, 23.0 to 230 ºF
Unit: ºC, ºF
Resolution: 0.1
Accuracy: ±0.2ºC
Measurement:
Reading Mode: Auto Read (Fast, Medium, Slow), Timed, Continuous
Reading Prompts: Reading, Stable, Locked
Temp. Compensation: ATC, MTC
Data management:
Data Storage: 500 results each
GLP Features: Yes
Inputs:
Conductivity with Temp. Probe: 5-pin aviation connector
Outputs:
USB: PC
RS 232: printer
Display options:
Backlight: Yes
Auto Shutdown: 1~60 min, off
IP Rating: IP54
Date and Time: Yes
General:
Power: AC Adapter, 100-240V AC input, DC9V output
Dimensions: 242×195×68 mm
Weight: 900g (1.98 lb)
Includes:
DJS-1VTC EC electrode
1413μS/cm EC standard solution, 50ml, 1 vial
REX-5 electrode stand