REX M510T(Optical DO) Multi-parameter Analyzer (20.000 pH, 2000.00 mV, 20.000) pX, 9.999e+9, 0.000 μS/cm ~ 3000 mS/cm)
Model: M510T
Parameters: pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation)
pH Measurement
Range: (-2.000~20.000) pH
Resolution: 0.1 pH, 0.01 pH, 0.001 pH
Accuracy: ±0.002 pH
Calibration Points: Up to 8
Standard Customization: Yes
Calibration Reminder: Yes
Standard Recognition: USA, NIST, MERK, JIS, GB, DIN, GOST buffers
Slope Limit: Yes
mV/ORP
Range: (-2000.00~2000.00) mV
Resolution: 1 mV, 0.1 mV, 0.01 mV
Accuracy: ±0.1 mV or ±0.03%
EH ORP Mode: Yes
ORP Calibration Points: 1 custom point (relative mV)
pX
Range: (-2.000~20.000) pX
Resolution: 0.1 pX, 0.01 pX, 0.001 pX
Accuracy: ±0.002 pX
Calibration Points: Up to 8
ISE (Ion Selective Electrode)
Range: 1.000e-9 ~ 9.999e+9
Units: mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb
Resolution: Up to 4 significant digits
Accuracy: ±0.3%
Calibration Points: Up to 8
Conductivity
Range: 0.000 μS/cm ~ 3000 mS/cm
Resolution: 0.001 μS/cm minimum, changed with range
Accuracy: ±0.5% FS
Reference Temperature: 5 ºC, 10 ºC, 15 ºC, 18 ºC, 20 ºC, 25 ºC
Calibration Points: Up to 5
Calibration Reminder: Yes
Standard Recognition: 10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm; 146.5 μS/cm, 1408 μS/cm, 12.85 mS/cm, 111.3 mS/cm
Resistivity
Range: 5.00 Ω·cm ~ 100.0 MΩ·cm
Resolution: 0.01 Ω·cm minimum, changed with range
Accuracy: ±0.5% FS
TDS
Range: 0.000 ppm ~ 1000 ppt
Resolution: 0.001 ppm minimum, changed with range
Accuracy: ±0.5% FS
Salinity
Range: (0.0~80.0) ppt
Resolution: 0.1 ppt
Accuracy: ±1 ppt
Dissolved Oxygen (Optional fluorescent DO electrode)
Sensor Type: DO-968-HC Optical DO electrode
Range: (0.00~20.00) ppm
Resolution: 0.01 ppm
Repeatability: 0.15 ppm
Accuracy at 0% DO: ≤0.1 ppm
Measurement Accuracy: ±0.30 ppm
Response time: ≤45s (90% response at 20 °C)
Salinity Compensation Accuracy: ±2%
Calibration Reminder: Yes
Calibration Points: Air-saturated water or zero point
Barometric Compensation: Yes
Manual Salinity Factor Correction: Yes
Saturation (Optional fluorescent DO electrode)
Range: (0.0~200.0)%
Resolution: 0.1%
Measurement Accuracy: ±10.0%
Dissolved Oxygen Concentration (Polarographic)
Sensor Type: DO-958-Q polarographic DO electrode
Range: (0.00~99.99) ppm
Resolution: 0.01 ppm
Accuracy:
±0.10 ppm, (0.00~50.00) ppm
±0.50 ppm, (50.0~99.99) ppm
Calibration Reminder: Yes
Calibration Points: Air-saturated water or zero point
Barometric Compensation: Yes
Manual Salinity Factor Correction: Yes
Saturation (Polarographic)
Range: (0.0~600.0)%
Resolution: 0.10%
Accuracy: ±2.0%
Temperature
Range: (-10.0~135.0) ºC, (14.0~275.0) ºF
Unit: ºC, ºF
Resolution: 0.1 ºC, 0.1 ºF
Accuracy: ±0.1 ºC
Measurement
Reading Mode: Auto Read (Fast, Medium, Slow), Timed, Continuous
Reading Prompts: Reading, Stable, Locked
Temp. Compensation: ATC, MTC
Data management
Data Storage: 1000 Groups
GLP Features: Yes
Inputs
pH Electrode: BNC (Q9)
DO with Temp. Probe: 4-pin aviation connector
Conductivity with Temp. Probe: 5-pin aviation connector
Outputs
USB: USB 2.0 flash drive, PC, Bluetooth Printer
Display options
Backlight: Yes
Auto Shutdown: 300, 600, 1200, 1800, 3600 sec., off
IP Rating: IP67
Date and Time: Yes
General
Power: Rechargeable Lithium battery, AC Adapter (Input: 100-240V AC; Output: 5V DC, 2A)
Dimensions: 90×255×40 mm
Net Weight: 500 g