1. Power supply:AC(220±10%)V,50Hz
2. Flash point determination:
Range: Ambient to 300℃ ; Repeatability: ≤3℃; Reproducibility:≤6℃; Accuracy:0.1℃
3. Heating rate:Procedure A: (5~6)℃/min, Procedure B: (1~1.6)℃/min
Automatic control and manually adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5. Igniting mode: Electric ignition. Gas flame diameter:3.2mm~4.8mm
6. Working condition: Ambient temperature:(10~40)℃
Relative humidity:≤80%
7. Maximum power consumption:500W
8. Dimension: 520mm×360mm×310mm