Average Power Measurement:
Frequency Range: 300-4200MHz
Power Range: 100mW-200W
Dynamic Range (dB): ≥33
Insertion Loss (dB): ≤0.1
VSWR: 1.05 to 99.9
Directivity: ≥30 (<3GHz); ≥28 (>3GHz)
Accuracy: ±4%+0.05W (0℃~+15℃ or +35℃~+50℃ increase 3%)
Impedance: 50Ω
Connector: N (Female)
Data Interface: USB
DPM Interface: DB9
Peak Power Measurement:
Peak Power Range: 100mW to 500W
Peak Power Accuracy: Burst width >200us: ±7%; 1us<Burst width< 200us: ±10%; 0.5us<Burst width< 1us: ±15%; Burst width< 0.5us: ±20%; NOTE: lower 700MHz or higher 3300MHz, increase +3% lower 15℃ or higher 35℃, increase +3%; Duty Cycle lower 0.1,increase +0.1W; Frequency lower 10HZ, +( 1.5%+0.15W)
Peak Average Ratio: 0 to12dB
CCDF:
Measurement Range: 0.1 to 100%
Measurement Accuracy: ±3%
Threshold Measurement Range: 0.05W to 500W
Burst Power:
Burst Power Range: 100mW to 200W
Burst Width: 1us to 60ms
Min. Measurement Frequency: 15HZ
Measurement Accuracy: ±6% +0.05W (0℃ to +15℃ or +35℃ to +50℃, increase +3%) Note:lower 700MHz or higher 3300MHz, increase +3%
Duty Cycle: 0.0001 to 1
General Specifications:
Power Supply (Monitor): NiMH Battery/ AC Adaptor
Operating Temperature: -10℃ to 50℃
Storage Temperature: -20℃ to 70℃
Relative Humidity: 0 to 85% (Non-condensing)
Weight: 480g (Sensor)/ 900g (Handheld monitor)
Dimensions (H×W×T): 130×(95)×124×34mm (Sensor); 220×110×70mm (Handheld monitor)
Ordering Information
Standard Package:
DPM sensor, USB cable, Software CD, Soft carry bag, Warranty card, CE certificate, Certificate of calibration, User’s manual.