인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Temperature Range: (-30~155)℃/ (-40~155)℃ customized (Ambient Temperature 25°C)
Temperature Deviation: ±0.15℃
Temperature Stabilization Time: 5 minutes
Display Resolution: 0.001℃
Insert Depth: 170 mm
Outside Diameter of Soaking Block: Ø25 mm
Standard Block Insertion Aperture: Ø6, Ø8, Ø10 mm or customized
Temperature Unit: ℃, ℉, K
Temperature Difference Between Holes: ≤±0.05℃
Axial Temperature Field: (Using special soaking block for temperature field test)
Temperature Fluctuation: Temperature change within 20mm from the bottom of the block insertion ≤ 0.5℃
Power Supply: 220VAC 50Hz 0.45 kW
Communication: RS232
Language: Chinese and English
Dimensions (Including Carrying Handle): (320 x 240 x 415) mm/ (420 x 240 x 415) mm
Weight:15.5kg/ 24.5 kg
Power: 0.45 kW