Input voltage: 220V AC±10%, 50Hz

Output voltage: 

DAC: 0-60 kV(peak)/0-42.4kV (rms)

Precision ±1%

Resolution 0.1kV

Frequency range 20Hz~500Hz

Capacity range: 1 nF ... 10 μF at 30 kVpeak; 1 nF ... 4.25 μF at 60 kVpeak

PD sensitivity range: 2 pC … 100 nC (acc. to IEC60270)

Resolution: ±1pC

System noise level: < 20 pC at 60 kV(peak)

PD localisation: 

Measuring range: 0 … 20,000m

Propagation velocity: 5 … 120m/µs

Sampling rate: 200 MHz

Bandwidth: 150kHz ~ 45MHz, automatic adjustment

Precision: 1% of the cable length

Resolution: ± 1pC / ± 0.1m

TD(Tan Delta) test: 

Measuring range: 0.1% ~ 10%, automatic adjustment

TDR joint localisation in: Integrated

calibration mode: 

Software: User selectable graphical interface, online “live” PD mapping integrated measurement database, comprehensive viewing processing, analysing and reporting of measurement data

Temperature: 

Operation: -20 °C … 55 °C

Storage: -30 °C … 70 °

Relative humidity: 93 % at 30 °C (non-condensing)

Weight: 100 kg(net)

Dimensions: Ø 690 x H 1000 mm