Relative Humidity: 3 % RH to 95 % RH range, accuracy ±2 % RH at 25 ºC
Typical Temp. Characteristics: -40 ºC to 150 ºC at 0.05 % RH/ºC
Operating Range: -40 ºC to +180 ºC
Time constant under 30 seconds
Temperature: -40 ºC to +180 ºC at ±0.5 ºC
Time constant under 4 seconds
RH and Temperature Outputs: 4 to 20 mA
Input Voltage: 7 to 30 VDC polarity protected
Dimensions:
Transmitter:
ABS housing, 120 mm x 80 mm (4.72 in x 3.14 in x 2.16 in)
Temp. operating range: -20 ºC to +70 ºC
Meets NEMA 4x, 13 specifications
Wall Probe: Stainless steel, 64 mm x 16 mm diameter (2.5 in x .625 in), includes metal wall mounting clip, 40 in. (1 M) teflon cable
Duct Probe:
Stainless steel, 216 mm x 16 mm diameter (8.5 in x .625 in),
1 M (40 in) teflon cable with adjustable or removable duct flange 70 mm (2.75 in) diameter
Probe Housings:
Stainless steel NEMA 4 enclosure
Connectors:
Liquid tight for .09 in. to .265 in. diameter cables. 4 pin plug-in screw terminal block for outputs. 5 pin screw terminal block for sensor cable.
Weight:
Wall probe with transmitter 397 grams (14 oz)