인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Resistance Ranges and Test Currents
Full Scale/ Resolution/ Accuracy/ Test Current
1.999 mΩ /1 μΩ /±(0.25% of reading + 1 count)/ 1.4 A
19.99 mΩ /10 μΩ/ ±(0.25% of reading + 1 count)/140 mA
199.9 mΩ/100 μΩ /±(0.25% of reading + 1 count)/ 14 mA
1.999 Ω / 1 mΩ/±(0.25% of reading + 1 count)/1.4 mA
19.99 Ω/ 10 mΩ/±(0.25% of reading + 1 count)/140 μA
Reactive loads not to exceed 2mH or 100µF with a minimum of 5µΩ per µH of series resistance.
Recommended Calibration Cycle: 1 year
Environmental
Operating Temperature: 0 to + 50˚ C
Storage Temperature: -40˚C to + 71˚ C
Physical and power
Size: 14.1 in W x 10.6 in D x 6.1 in H
Weight 9.8 lb (4.45 kg)
Power: 90 to 250 VAC 50/60 Hz
3 High Capacity ‘C’ NiCad rechargeable batteries are included
*Accuracy based on recommended calibration cycle.