TEST CURRENTS: 1 mA, 10 mA, 100 mA, 1 A, 10 A, 100 A. Each current may be continuously adjustable from 0 to 100%.
RESISTANCE RANGES:
Current Range Resolution
100 A 0-199.9 µΩ 0.1 µΩ
100 A 0-1999 µΩ 1 µΩ
10 A 0-1999 µΩ 1 µΩ
10 A 0-19.999 mΩ 1 µΩ
1 A 0-19.99 mΩ 10 µΩ
1 A 0-199.99 mΩ 10 µΩ
100 mA 0-199.9 mΩ 100 µΩ
100 mA 0-1999.9 mΩ 100 µΩ
10 mA 0-1999 mΩ 1 mΩ
10 mA 0-19.999 Ω 1 mΩ
1 mA 0-19.99 Ω 10 mΩ
1 mA 0-199.99 Ω 10 mΩ
For each test current, ranges are automatically selected for optimal reading.
MEASUREMENT PRINCIPLE: Four-terminal, Kelvin-type.
CONTINUOUS OPERATION TIME: At 100 A this equipment may be used continuously for aprox. 15 minutes before the thermal protection activates.
At 10 A or less, there is not a limited time for continuous operation.
BASIC ACCURACY:
R < 0.5 mΩ: ± (0.50% of reading + 2 ULSD*).
R ≥ 0.5 mΩ: ± (0.20% of reading + 2 ULSD*).
* Units of the Least Significant Digit.
ADVANCED FEATURES: Digital direct reading of very low resistances in the alphanumerical display, with up to 4½ digits. Very fast and accurate measurements.
TERMAL PROTECTION: Protects all sensitive components, avoiding any damage due to overheating.
ENVIRONMENTAL: IP54 with closed lid.
SAFETY CLASS: Meets the requirements of IEC 61010-1.
INTERFACE: USB.
POWER SUPPLY: Mains or internal battery powered.Internal battery: rechargeable, sealed lead-acid (for up to 10 A test current).
Mains supply*: 100-130 V~ or 220-240 V~ (for up to 100 A test current).
*The option must be indicated on the order.
BUILT-IN BATTERY CHARGER: For 100-130 V~ or 220-240 V~ mains supply.
OPERATING TEMPERATURE RANGE: -5 °C to +50 °C.
STORAGE TEMPERATURE RANGE: -25 °C to +70 °C.
HUMIDITY RANGE: 95% UR (non condensing).
EQUIPMENT WEIGHT: Approx. 14 kg.
DIMENSION: 502 x 394 x 190 mm.
AVAILABLE ACCESSORIES: 2 Combined current and potential leads for 10 A. 2 Combined current and potential leads for 100 A. 1 Power cord. 1 USB cable. 1 User guide. 1 Case for the accessories.