인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결

Withstand Voltage Test:
Output Voltage:
+ AC: 0.05-5kV, Load Variance: 1%, Accuracy: 1%, Resolution: 2V
+ DC: 0.05-6kV, Load Variance: 1%, Accuracy: 1%, Resolution: 2V
Current Test Range:
+ AC: 0.001mA - 40mA, Accuracy:1%, Resolution: 1μA
+ DC: 0.0001mA - 20mA, Accuracy: 1%, Resolution: 0.1 μA
Output Power: 200VA
Insulation Resistance Test:
Output Voltage: 0.05 - 5kV, Resolution: 2V, Accuracy: 1% of set value + 0.1% full scale
Resistance Test Range: 1MΩ-50.0GΩ Resolution:0.1MΩ
Discharge Function: Automatic discharge after the end of the test
ARC Detection:
+ AC: 1mA - 20mA
+ DC: 1mA - 10mA
Contact Check Function: OSC open and short:600Hz, 0.1s
Security Features:
High Voltage Floating Output: ---
Electric Shock Protection: 0.5mA ±0.25mA
Other Protection: Start protection, panel operation password protection
Alarm Indication: PASS: short tone, green light; FAIL: long tone, red light
Memory: groups, 50 steps per group
General Parameters:
Voltage Rise Time: 0.1s — 999.9s
Test Time Setting(AC/DC): 0.3s — 999s
Voltage Fall Time: 0.1s — 999.9s
Waiting Time (IR): 0.2s — 999.9s
Time Accuracy: ±(1%+0.1s)