Acceleration Input
Impedance: 100 kΩ or greater
Sensitivity: Current On: 5 or 10 mV/g
Sensitivity: Current Off: 1 or 10 mV/g
Max Input: ±1000 g
Accuracy
Acceleration (at 100 Hz REF): 10, 100, or 1000 range ±1%
Frequency Response (from REF): Acceleration 3 Hz to 20 kHz ±3%; output is 3 dB down at selected filter cutoff
Current: 3.5 ±1 mA
Connector: BNC
AC Output
Impedance: <100 Ω
Sensitivity: 5 Vrms or 5 V FS
Frequency Response: 3 Hz to 20 kHz
Connector: BNC
Interface – Power
Power: 6 V batteries; nominally 8-hr operation before charging
Controls
Meter Sensor mV/g Switch: 4-position switch selects “5” or “10” mV/g with current on or “1” or “10” mV/g with current off
Meter FS Power Switch: Selects “10,” “100,” “1000” FS range or “Power Off”
Filter Switch (3 dB cutoff 10%):
Position 1: 3 Hz HP, 2 kHz LP
Position 2: 3 Hz HP, 5000 Hz LP
Position 3: 3 Hz HP, 10,000 Hz LP
Position 20: 3 Hz HP, 20,000 Hz LP
Indicators
Vibration Meter: 3.5 digit DVM with LCD display
Sensor Open LED: Illuminates when the sensor is open
Sensor Short LED: Illuminates when the sensor is shorted
Batt LO LED: Illuminates when battery needs to be charged
OL LED: Illuminates if the input is overloaded or any FS is exceeded
Mechanical
Dimensions: 6" W × 7" H × 2.5" D (15.2 cm × 17.8 cm × 6.3 cm)