인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
- Range and accuracy:
Logger operating range: 0 ... +50 °C (32 ... +122 °F)
Probe operating range: −95 ... +70 °C (−139 ... +158 °F)
Calibrated measurement range: −55 ... +50 °C (−130 ... +122 °F)
Resolution: 0.01 °C at +25 °C (0.02 °F at +77 °F)
Accuracy over temperature range at −55 … +50 °C (−67 … +122 °F): ±0.5 °C (±0.9 °F)
- Memory:
Memory type: Non-volatile EEPROM
Memory mode: User-selectable wrap (FIFO) or stop when memory is full
- Data sample capacity:
DL1016MR/MRB: 68 600 16-bit samples
- Dimensions: 85 × 59 × 26 mm (3.4 × 2.3 × 1 in)
- Weight: 76 g (2.7 oz)
- Internal sensor type: Precision-tolerance epoxy encapsulated NTC thermistor
- Cable construction: 2 mm (0.07 in) diameter, Teflon coated cable
- External temperature probes:
Sensor tip: Stainless steel
Diameter: 3.2 mm (1/8 in)
Length: 38 mm (1.5 in)
- Probe cable lengths:
DL1016MR/MRB: 3 m (10 ft)