Measurement range: 0 … 1 aw
T90 response time: 10 min
Sensor: HUMICAP 180L2
Accuracy: ±0.01 aw (±1 %RS)
Factory calibration uncertainty
±0.6 %RH (0 … 40 %RH)
±1.0 %RH (40 … 95 %RH)
Temperature
Measurement range: −40 … +180 °C (−40 … +356 °F)
Accuracy at +20 °C (+68 °F): ±0.2 °C (0.36 °F)
Moisture in transformer oil ppm accuracy: ±10 % of the reading