Input Measurement Function
DCV
Range Measurement Range Resolution Accuracy
100 mV −10.0 mV ~ 110.0 mV 100 µV 0.2% + 4
30 V −5.00 V ~ 30.00 V 10 mV 0.2% + 4
DCA
Range Measurement Range Resolution Accuracy
20 mA 0.00 mA ~ 22.00 mA 0.01 mA 0.2% + 4
LOOP
Range Measurement Range Resolution Accuracy
24 V / ±10% /
On / off
Range Measurement Range Resolution Accuracy
/ OPEN / CLOSE / /
Notes
★ DCV characteristics: Input impedance: 30V: 300kΩ; mV range: >1MΩ
★ DCI characteristics: Overload protection: 50mA / 30V; Load voltage: approximately 18mV/mA
★ TC description: The thermocouple measurement adopts the ITS-90 temperature scale. The accuracy does not include the error of cold junction compensation and the influence of thermoelectric potential.
Analog Output Function
DCV
Range Measurement Range Resolution Accuracy
100 mV 0.0 mV ~ 110.0 mV 100 µV 0.2% + 4
20 V 0.00 V ~ 24.00 V 10 mV 0.2% + 4
DCI (Active)
Range Measurement Range Resolution Accuracy
20 mA 0.00 mA ~ 22.00 mA 0.01 mA 0.2% + 4
DCI (Passive)
Range Measurement Range Resolution Accuracy
−20 mA 0.00 ~ −22.00 mA 0.01 mA 0.2% + 4
★ DCI feature: 20 mA maximum load 1 kΩ
★ FREQ characteristics: Rectangular wave, 60% duty cycle, 1 to 20 Vp-p
★ TC description: The thermocouple measurement adopts the ITS-90 temperature scale. The accuracy does not include the error of cold junction compensation and the influence of thermoelectric potential.