인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
DC Output Voltage: 60V to 11000V
Accuracy: +/- (0.25% rdg + 0.2V)
Output Resolution: 1V
Loading: Up to 4mA charging, 2mA continuous
Current Measurement:
●Returns current only, from 1pA to 4mA
●Accuracy (Resistive Load up to 1000pF): +/- (0.5%rdg + 10pA + 2pA per kV))
●Accuracy (Capacitive Load 1000pF to 50nF): +/- (0.5%rdg + 100pA + 20pA per kV))
●Accuracy (Capacitive Load >50nF): +/- (0.5%rdg + 1nA + 200pA per kV))
Specifications: Valid after 15-minute warmup, for 1 year from last external calibration, and for ambient temperature within +/-2°C of last performed ZERO operation.
Ethernet: High-speed, high-noise immunity LAN interface
RS232 Interface: Selectable baud: 9600, 19200, 38400, 57600, or 115200; full handshake
VICL Interfaces: Provided for control of HV Scanners
Digital I/O Interface: Provides 4 digital inputs and 5 digital outputs. Functions include Start/Stop, Testing, Pass/Fail, HV Present, Safety Interlock, Dwell Timer
Optional GPIB: Option GPIB-9 adds GPIB interface capability
Operating Temperature: 0°C to 50°C
Humidity: <85% RH (non-condensing)
Power: 105-265Vrms, 45-65Hz, 500VA Max
Dimensions: 89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D)
Weight: 5Kg (11 lb.) Net / 7Kg (15 lb.)
Accessories: Alligator test leads, NIST traceable calibration certificate with no data, power cord, and operator's manual. Limited ISO 17025 cal cert with data & uncertainties available.