인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Operating principle: Ultrasonic, pulse echo measurement method
Measuring range: Depends on probe and material, 0.60 mm to 508 mm (0.025” to 20.00”) in IP to 1st BW measurement mode, 2.00 mm to 127.0 mm (0.079” to 5.00”) in dual-multi measurement mode, the coating thickness range shall be 0.3 mm to 2.50 mm (0” to 0.098”).
Measuring resolution: 0.01 mm default - selectable 0.01, 0.1 mm (0.001” default - selectable 0.001”, 0.01”)
Material velocity range: 0.508 to 18.699 mm/msec (0.0200” to 0.7362”/msec)
Material velocity resolution: 1 m/s (0.0001”/msec)
Units: Inch or millimeter
Calibration: One-point cal, on-block and off-block, two-point cal
Pulser
Excitation Pulse: Spike pulser
Voltage: 120 V into 50 ohm load, using 20 MHz oscilloscope
Receiver
Bandwidth: 500 KHz to 12 MHz @ -3 dB
Gain: Automatic gain control
Display type: High-resolution graphical LCD, 64 x 128 pixels, 53.0 mm x 27.0 mm with backlight and adjustable contrast
Update rate: 4 Hz or 8 Hz, user selectable, 24 Hz scan mode capture rate
Thickness value display
NORMAL view mode: 5 Digit, 10.6 mm (0.4”) high
B-SCAN view mode: 5 Digit, 2.55 mm (0.1”) high
Display of last reading: Solid filled or hollow digits indicate coupled or uncoupled condition
Setups: 9 standard setups for probes
Alarm settings: Minimum and maximum alarms, range of 0.25 mm to 508 mm, 0. (0.010” to 20.00”) reading alternates between solid and hollow when alarms are enabled and violated
Power requirements: 2 “AA” size batteries
Battery life/operating time: Approximately 60 hours
Instrument shut-off Selectable ALWAYS ON or AUTO OFF after 5, 10, 15, 30 minutes of inactivity
Language: Selectable English, German, French, Spanish, Italian, Russian, Japanese and Chinese
I/O connectors
Transducer: Dual Lemo 00 (coax)
Mini-USB: Mini USB to PC
Temperature
Operating -10°C to +50°C (+10°F to +120°F)
Storage -20°C to +60°C (-10°F to +140°F)
Weight 223 g (0.597 lb) including batteries
Size 138 mm x 32 mm x 75 mm
Shock IEC 68-2-27 Ea, as per Mil Std 810C Method 516.2 Procedure I with a 15g 11ms impulse half sinusoidal wave applied 6 times per axis
Sealing IEC529 / IP54, dust proof/dripping water proof as per IEC 529 specifications for IP54 classification
Data recorder option features
Capacity: 50,000 readings
File structures: Grid file
Number of rows: 1 to 50,000
Number of columns: 1 to 223
File naming: Up to 24-character alphanumeric name
Optional software: UltraMATE and UltraMATE Lite