ESCO CCL-170B-8-NF-HHS CelCulture® CO₂ 인큐베이터 (170 L, 180°C HHS, 230 VAC 50/60 Hz)
생산업체: ESCO Model: CCL-170B-8-NF-HHS P/N: 2170298 - 연락처
IR SENSOR MODEL WITH STAINLESS STEEL CHAMBER (NO ULPA FILTER)
CelCulture® Incubator 170 L IR Sensor, CO2 Control, 180°C HHS, 230 VAC 50/60 Hz, No ULPA Filter
TEMPERATURE:
Temperature Control Method: Direct Heat and Air Jacket using Microcontroller PI
Ambient Temperature Range: 18 to 32°C (64.4 to 89.6°F)
Temperature Control Range, °C: Ambient +5 to 60
Temperature Uniformity, °C: Standard Unit: <± 0.35; Suppressed O2 model: <± 0.4
Temperature Fluctuation, °C: ±0.2
Temperature Recovery Time (after 30 seconds door opening, 98% from initial value): ≤5 minutes
CO2:
CO2 Control System: Microcontroller PI
CO2 Control Range, % CO2: 0-19.5
CO2 Fluctuation, % CO2: ± 0.2
CO2 Sensor: Infrared (IR) Sensor
CO2 Recovery Time (after 30 seconds door opening, 98% from initial value): At 5.0% CO2 by volume (Standard unit): ≤5 minutes; Suppressed O2 model: ≤8 mins.
HUMIDITY:
Humidification Method: Humidity pan
Humidity Range, %RH (at 37°C): 85-90
PHYSICAL CONSTRUCTION:
Interior Volume: 170 L
External Dimensions (W x D x H): 660 x 660 x 906 mm
Internal Dimensions (W x D x H): 505 x 535 x 633 mm
Net Weight: 101 kg
Chamber Construction: Main Body Electrogalvanized steel with Isocide™ antimicrobial coating; Interior Material Stainless steel, type 304
Number of Shelves: 4
Maximum Number of Shelves: 7
Electrical Configuration 110-130 VAC, 50/60 Hz
Full Load Amps: 10 A
Electrical Configuration 220-240 VAC, 50/60 Hz
Full Load Amps: 5 A
CONTAMINATION CONTROL
Contamination Control Methods
1) Main body is electrogalvanized steel with Isocide™ antimicrobial coating;
2) 180˚C high heat sterilization cycle;
3) ULPA filter (optional) - filter must be removed during sterilization cycle
4) 0.2 µm gas inlet filter
5) 1-micron air circulation filter
All data recorded were observed with unloaded chambers and under optimum factory setting of 22 ±3°C with room humidity of 30-60%.
*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.
**For temperature not exceeding 37.2°C.
***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.
****For CO2 level not exceeding 5.2%.
*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.
****** For O2 level not lower than 4.8%.
******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%